Two Position Injector and Valve Applications

Application guide

Two Position Injector and Valve Applications

These illustrations show basic sample injection techniques using Valco two position valves. With rare exceptions, there is no difference between switching valves and external volume sampling valves, so the same valve can be used for either function.

The unique advantage of 8 and 10 port valves is that they reduce extra column volume by combining sampling and switching functions in a single valve. This minimizes expense, maintenance, service, and risk of leaks as compared to multiple 6 port valve systems.

Explore the flow paths below. Open an application to watch the valve switch positions automatically. Use Pause to study a diagram, or choose Position A or Position B. Reduced-motion preferences start playback paused.

Browse all 18 applications

Alternate column regeneration

10 port

When columns must be regenerated following each analysis, this technique permits automation of the process. While one column performs the analysis, the second column undergoes regeneration by use of an auxiliary pump. Once the first analysis is complete, the valve is switched and the regenerated column is now ready for analytical use.

Watch the video on YouTube

10-port valve, Position A: Column 1 performing separation, column 2 being regenerated
Position A: Column 1 performing separation, column 2 being regenerated
10-port valve, Position B: Column 1 being regenerated, column 2 performing separation
Position B: Column 1 being regenerated, column 2 performing separation

Backflush of precolumn to vent

6 port

This plumbing scheme allows slower eluting components (end cut) which are not of interest to be backflushed to vent. Often a shorter version of the analytical column is used as the precolumn.

Once all components of interest have entered the main column (at port 2), the valve is switched, backflushing the precolumn to vent and reducing analysis time. An auxiliary source of carrier or mobile phase is required for this application

Watch the video on YouTube

6-port valve, Position A: Slower-eluting compounds collected in precolumn
Position A: Slower-eluting compounds collected in precolumn
6-port valve, Position B: Contents of precolumn backflushed to vent
Position B: Contents of precolumn backflushed to vent

Detector selection from two columns, or one column and auxiliary carrier

4 port

This unique configuration allows analysis of different parts of one analysis with two different detectors, without splitting or multiple injections. For example, fixed gases can be analyzed with a thermal conductivity detector, followed by the analysis of a hydrocarbon fraction with a flame ionization detector.

Watch the video on YouTube

4-port valve, Position A: Contents of column to Detector 1
Position A: Contents of column to Detector 1
4-port valve, Position B: Contents of column to Detector 2
Position B: Contents of column to Detector 2

Dual microvolume sample injection

6 port

The CI6 injector shown in this application is not on this website. Please contact us for details.

This microvolume injector can be used to alternate between two different samples. Each time the valve is switched, a sample is injected.

When the two sample inlets are connected in series, the valve injects the sample each time the valve switches. This is particularly useful in heavy duty cycle operations to minimize valve wear.

When the sample/waste and pump/column connections are swapped, the valve can be used to inject the same sample onto two different columns, in alternation.

Watch the video on YouTube

6-port valve, Position A: Sample 1 to column
Position A: Sample 1 to column
6-port valve, Position B: Sample 2 to column
Position B: Sample 2 to column

Heart cut trapped in a loop and injected onto a second column

10 port

Sample is injected (using a separate injector) onto an analytical column. Early eluting components (front cut) pass through a trapping loop and are detected at port 3. The valve is then switched, and the center (or heart cut), which was retained in the trapping loop, is injected onto the second column to the detector (at port 4).

Late eluting components (end cut) are trapped on the first column. When the valve is switched again, the end cut passes through the trapping loop to the first detector, completing the analysis.

Watch the video on YouTube

10-port valve, Position A: Early eluting components go to Detector 1
Position A: Early eluting components go to Detector 1
10-port valve, Position B: Heart cut to Detector 2
Position B: Heart cut to Detector 2

Loop sampling with backflush of precolumn to vent

10 port

When components of interest are low boiling, this plumbing scheme allows "heavy" components with long retention times to be backflushed to waste. After the sample loop is loaded in Position A, the valve is switched to Position B to inject the sample onto column 1. As soon as all components of interest have entered column 2, the valve is switched back to Position A. Column 1 is backflushed to vent during the analysis, reducing the total analysis time.

Watch the video on YouTube

10-port valve, Position A: Sample loop loaded/Column 1 backflushed
Position A: Sample loop loaded/Column 1 backflushed
10-port valve, Position B: Sample to columns
Position B: Sample to columns

Loop sampling with backflush to detector

8 port

One valve performs the functions of sampling and backflush valves, simplifying operation and reducing cost. When components of interest are detected, the strongly retained components are backflushed and removed from the column, without temperature programming.

Watch the video on YouTube

8-port valve, Position A: Contents of sample loop injected onto column
Position A: Contents of sample loop injected onto column
8-port valve, Position B: Late eluting components backflushed to detector
Position B: Late eluting components backflushed to detector

Loop sampling with two column sequence reversal

10 port

This is ideal for fixed gas-from-CO2 analysis where no "high boilers" are present. Column 1 is packed with a porous polymer and Column 2 with molecular sieve. The sample loop is loaded in Position A. When the valve is switched, the loop contents are sent onto Column 1.

As the inorganic gases and methane leave Column 1 and enter Column 2, the valve is returned to Position A, reversing the column sequence. CO2 now leaves Column 1, becoming the first peak. The inorganics and methane are separated by the mole sieve and pass through the porous polymer column to the detector.

Watch the video on YouTube

10-port valve, Position A: Sample loop loaded/Column sequence reversed
Position A: Sample loop loaded/Column sequence reversed
10-port valve, Position B: Loop contents sent to Column 1
Position B: Loop contents sent to Column 1

Microvolume sample injection

4 port

The internal sample (fixed volume) flowpath is used when very small volumes are required. The sample size is determined by a passage engraved on the valve rotor, allowing precise, repeatable injections.

In Position A, the sample flows through the sample passage while the mobile phase flows through to the column. The third passage is inactive. In Position B, the sample passage is in line with the column and the mobile phase injects the contents of the sample passage onto the column. The passage which was inactive in Position A allows the sample to continue flowing without interruption.

Watch the video on YouTube

4-port valve, Position A: Fill internal sample slot
Position A: Fill internal sample slot
4-port valve, Position B: Inject contents of internal sample slot
Position B: Inject contents of internal sample slot

Same sample to different loops

8 port

In a dual external sample loop configuration, sample is injected in both positions. In Position A, Loop 2 is loaded while the mobile phase flows through Loop 1 and onto the column. In Position B, the Loop 2 sample is injected onto the column and another sample is loaded into Loop 1. When the valve is returned to Position A, the Loop 1 sample is injected onto the column and Loop 2 is reloaded.

Watch the video on YouTube

8-port valve, Position A: Loop 1 injected onto column, Loop 2 loaded
Position A: Loop 1 injected onto column, Loop 2 loaded
8-port valve, Position B: Loop 2 injected onto column, Loop 1 reloaded
Position B: Loop 2 injected onto column, Loop 1 reloaded

Sample enrichment (cleanup) using dual precolumns

10 port

Sample is injected by a separate injector onto one of two precolumns (stripper). Early eluting components vent at port 6 while components of interest are retained on the stripper. When the valve is switched, a new injection is made onto the second stripper while components retained on the first stripper are backflushed onto the analytical column at port 9. This method requires an auxiliary pump at port 4.

Watch the video on YouTube

10-port valve, Position A: Early-eluting components are vented, components of interest retained on precolumn 1
Position A: Early-eluting components are vented, components of interest retained on precolumn 1
10-port valve, Position B: Components of interest sent to analytical column, new injection repeats the process with precolumn 2
Position B: Components of interest sent to analytical column, new injection repeats the process with precolumn 2

Sample injection

6 port

With the valve in Position A, sample flows through the external loop while the carrier flows directly through to the chromatographic column. When the valve is switched to Position B, the sample contained in the sample loop and valve flow passage is injected onto the column.

Watch the video on YouTube

6-port valve, Position A: Sample loop is filled
Position A: Sample loop is filled
6-port valve, Position B: Contents of the loop are injected onto the column
Position B: Contents of the loop are injected onto the column

Sample injection and heart cut to detector

10 port

The sample loop is loaded in Position A. When the valve is switched, the contents of the loop are injected onto the analytical column. The front cut elutes from the column and goes to vent at port 5. The valve is then switched back to Position A, and the heart cut is backflushed to the detector). Once the heart cut has eluted, the valve is switched back to Position B, sending the end cut to waste.

By delaying the switching times, the front and heart cuts could go to waste, with the end cut going to the detector.

Watch the video on YouTube

10-port valve, Position A: Sample loop loaded/Heart cut to detector
Position A: Sample loop loaded/Heart cut to detector
10-port valve, Position B: Contents of loop to column/End cut to waste
Position B: Contents of loop to column/End cut to waste

Simultaneous injection of the same sample onto two separate columns

10 port

In Position A, sample fills the two loops in series. In Position B, the sample is simultaneously injected into two separate flow systems. A single autosampler used with this flowpath can automate two analytical procedures for the same sample.

In an important non-chromatographic application, the roles of carrier and sample are reversed, permitting two different quantities of two different materials to be dispensed together, as in automatic dilution.

Watch the video on YouTube

10-port valve, Position A: Loops 1 and 2 filled in series
Position A: Loops 1 and 2 filled in series
10-port valve, Position B: Contents of Sample loop 1 injected onto Column 1, contents of Sample loop 2 injected onto Column 2
Position B: Contents of Sample loop 1 injected onto Column 1, contents of Sample loop 2 injected onto Column 2

Simultaneous injection of the same sample onto three separate columns

14 port

The 14 port valve permits a variety of plumbing arrangements. This one loads the same sample onto three sample loops for injection onto three different columns. This is used primarily by column manufacturers for QA since samples are injected simultaneously. It has also seen use in labs to perform different parts of the same analysis.

Watch the video on YouTube

14-port valve, Position A: Three sample loops loaded in series
Position A: Three sample loops loaded in series
14-port valve, Position B: Contents of three loops injected on three columns
Position B: Contents of three loops injected on three columns

Two column selection

6 port

When two different columns are required at frequent intervals at similar oven temperatures, a 6 port valve can provide rapid selection of the one to be used. The column not in use is protected by a blanket of inert mobile phase, and may be rapidly brought to equilibrium when required. (An 8 or 10 port valve is required if flow must be maintained to the non-selected column.)

Watch the video on YouTube

6-port valve, Position A: Column 1 selected
Position A: Column 1 selected
6-port valve, Position B: Column 2 selected
Position B: Column 2 selected

Two different samples to same column

10 port

A single 10 port valve permits alternate injections from two loops, which may be identical or of different sizes. This technique replaces a 4 port switching valve (used as the sample selector) and a 6 port sample injector.

In Position A, Loop 2 is loaded with sample 2 while the mobile phase flows through Loop 1 and onto the column. In Position B, the Loop 2 sample is injected onto the column and Loop 1 is loaded with sample 1. When the valve is returned to Position A, the Loop 1 sample is injected onto the column and Loop 2 is reloaded with sample 2.

Watch the video on YouTube

10-port valve, Position A: Contents of Sample loop 1 injected, Sample loop 2 loaded
Position A: Contents of Sample loop 1 injected, Sample loop 2 loaded
10-port valve, Position B: Contents of Sample loop 2 injected, Sample loop 1 loaded
Position B: Contents of Sample loop 2 injected, Sample loop 1 loaded

Two different samples to two columns

12 port

This 12 port plumbing arrangement is simply a dual 6 port diagram; while sample 1 is loaded and subsequently injected onto column 1, sample 2 is loaded and injected onto column 2. As with 10 port valves, the 12 port is capable of a wide variety of sample injection and switching applications.

12-port valve, Position A: Sample loop 1 is filled with sample 1 while sample loop 2 fills with sample 2.
Position A: Sample loop 1 is filled with sample 1 while sample loop 2 fills with sample 2.
12-port valve, Position B: Sample loop 1 contents are sent to column 1 and detector 1; sample loop 2 contents are sent to column 2 and detector 2. When the valve is returned to Position A, the loops are refilled.
Position B: Sample loop 1 contents are sent to column 1 and detector 1; sample loop 2 contents are sent to column 2 and detector 2. When the valve is returned to Position A, the loops are refilled.

Blank valve templates

Use these blank diagrams to sketch your own valve plumbing.

Find the right valve for your application

Our technical team can help with valve selection and plumbing.